ProntoPress-20
Instruction Manual
23
1. Advanced Operations
A Recirculation Cooling Unit (TRECI) can be connected by means
of a Recirculation Cooling Unit Connector (ROPRE).
Connect the cable from the pump in TRECI to ROPRE, according
to the diagram of ROPRE.
Connect the cable with plug from ROPRE to the socket
(pos. ) on the back of the mounting press.
Connect the 3-phase cable from ROPRE to the mains.
Check that the pump rotates in the direction indicated on the
pump. If the direction is incorrect, switch two of the phases in
the connection to the pump in TRECI (3-phase versions only).
Replace the outlet tube on the pump in TRECI with the non-
return valve with elbow and hexagon nipple from ROPRE.
Struers recommends that you seal the threads.
Connect the pressure hose of the mounting press to the
hexagon nipple. Remember to mount the gasket.
Lead the outlet hose (pos. ) to the inlet on TRECI. Be
absolutely sure to place the hose with a steady downward
slope and without obstructions.
Attach the water outlet to the inlet of TRECI.
Installation of Recirculation
Cooling Unit (Optional)
Socket for Recirculation
TRECI
LaboPress-3/
ProntoPress-10/ProntoPress-20
ROPRE
Outlet hose
Socket for recirculation
➀
➃
➄
➁
Summary of Contents for ProntoPress-10
Page 1: ...Manual No 15077003 Date of Release 11 01 2008 ProntoPress 20 Instruction Manual ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 60: ...ProntoPress 20 Gebrauchsanweisung Handbuch Nr 15077003 Auslieferungsdatum 11 01 2008 ...
Page 61: ......
Page 63: ......
Page 65: ......
Page 121: ...ProntoPress 20 Mode d emploi Mode d emploi no 15077003 Date de parution 11 01 2008 ...
Page 122: ......
Page 124: ......
Page 126: ......
Page 187: ......
Page 188: ......
Page 189: ......
Page 190: ......
Page 191: ......
Page 193: ......