![Struers ProntoPress-10 Instruction Manual Download Page 17](http://html1.mh-extra.com/html/struers/prontopress-10/prontopress-10_instruction-manual_1385652017.webp)
ProntoPress-20
Instruction Manual
11
The cylinder diameter has to be set, in order to have the right
values of force or pressure. When a cylinder diameter is changed,
the database values of force or pressure in the methods will
automatically change to maintain the conditions in the mounting
cylinder.
Note:
The counter which warns the user to clean the rams after
200 mountings, will be reset when a new cylinder diameter is
selected. It is therefore important to clean away any old cured
resin beneath the lower ram when changing cylinder diameter.
Please see the instructions in the ‘Maintenance’ section.
mO
mN
Press MENU UP/DOWN
mON
to select the Cylinder
parameters.
D
CONFIGURATION
Cylinder Diameter: L:
25 mm
R: 50 mm
Temperature Unit: CELSIUS
Force or Pressure: FORCE Unit: kN
D
e
Press ENTER
e
to activate the Cylinder Diameter
Menu.
D
CYLINDER DIAMETER
25 mm
30 mm 40 mm 50 mm
1 1/4" 1 1/2"
D
mO
mN
Press MENU UP/DOWN
mON
to select the diameter
for the cylinder, which is mounted.
D
e
Press ENTER
e
to accept the diameter.
D
Esc
Press Esc to return to the Select Method Group Menu.
Setting the Temperature Unit, Force or Pressure Mode, Force
Unit, Pressure Unit and Countdown Mode: See the section
Advanced Operations.
Setting the Cylinder Diameter
Summary of Contents for ProntoPress-10
Page 1: ...Manual No 15077003 Date of Release 11 01 2008 ProntoPress 20 Instruction Manual ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 60: ...ProntoPress 20 Gebrauchsanweisung Handbuch Nr 15077003 Auslieferungsdatum 11 01 2008 ...
Page 61: ......
Page 63: ......
Page 65: ......
Page 121: ...ProntoPress 20 Mode d emploi Mode d emploi no 15077003 Date de parution 11 01 2008 ...
Page 122: ......
Page 124: ......
Page 126: ......
Page 187: ......
Page 188: ......
Page 189: ......
Page 190: ......
Page 191: ......
Page 193: ......