47
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Information on
electromagnetic
compatibility (EMC)
Table 5
Recommended separation distances between portable and mobile RF communications equipment and the IMAGE1 S™ 4U Rubina,
OPAL1
®
NIR/ICG camera head
The IMAGE1 S™ 4U Rubina, OPAL1
®
NIR/ICG camera head is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or user of the IMAGE1 S™ 4U Rubina, OPAL1
®
NIR/ICG camera head can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the IMAGE1 S™ 4U Rubina, OPAL1
®
NIR/ICG camera head as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
[W]
Separation distance d [m] as a function of the transmitter frequency
150 kHz to 80 MHz
f
80 MHz to 800 MHz
f
800 MHz to 2.5 GHz
h
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of structures,
objects and people.
HF surgical instrument compatibility
This equipment has been evaluated for
compatibility with highfrequency surgical
equipment. It has been tested according to
IEC 6060122 Annex BB.