54
Table 204
G
uidance and manufacturer
’s
declaration – electromagnetic immunity –
for equipment and
s
y
s
tem
s
that are not life-
s
upporting
AIDA INTERFACE II model
20
0921 32 is intended for use in the electromagnetic environment specified below.
The customer or user of the AIDA INTERFACE II model
20
0921 32 should ensure that it is used in such an environment.
Immunity te
s
t
EN/IEC 60601
te
s
t level
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications e
q
uipment should be used
no closer to any part of the AIDA INTERFACE model
20
0911 32,
including cables, than the recommended separation distance calcu-
lated from the e
q
uation applicable to the fre
q
uency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to < 80 MHz
3
V
rms
>
@
P
d
3
/
5
.
3
150 kHz to < 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3
V/m
>
@
P
d
3
/
5
.
3
80 MHz to < 800 MHz
>
@
P
d
3
/
7
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance
level in each fre
q
uency range
b
.
Interference may occur in the vicinity of e
q
uipment marked with the
following symbol:
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects and
people.
a
Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the unit is used exceeds the
applicable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the unit.
b
Over the fre
q
uency range 150 kHz to 80 MHz, field strengths should be less than
3
V/m.
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit
Appendix
Electromagnetic Compatibility
(EMC) Information
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)