Connection to the Monitor 8403 ZX
The Flexible Intubation Video Endoscope can be used together with the KARL STORZ MONITOR 8403 ZX.
Warning:
It must be ensured that the electrically non-insulated parts of the Flexible Intubation Video Endoscope do
not come into contact with conductive surfaces, or voltage-carrying elements of other units, at any time.
Insert the connecting cord
q
of the Flexible Intubation Video Endoscope into either connection socket
s
or
d
on the
monitor.
The connecting cord
q
can be connected or disconnected during operation. Insert the cable from the power supply
into the connection socket
f
of the monitor and connect the power supply with the mains socket. The power cord
can be fixed to the connection using the union nut.
The battery is charged automatically as soon as the monitor is connected to the power supply. This will be indicated
by the lighting up of the orange colored LED
;
.
CAUTION:
The Flexible Intubation Video Endoscope can only be connected to the monitor 8403 ZX.
Use of the HDMI Output
The device features a port for an external monitor. An HDMI cable can be used to send the video signal to an
external monitor.
CAUTION:
The C-MAC
®
Monitor must be switched off when an external monitor is connected.
Note:
The connected monitor must support a resolution of 1024 × 768 or 1280 × 720.
Note
: To connect the C-MAC
®
Monitor to a KARL STORZ monitor, the HDMI-DVI adaptor with article number
20 9190 10 is required.
Installation and Operating
Instructions
20
f
d
s
d
f