Troubleshooting
Symptom
Possible causes
Remedy
Image unclear
Distal objective lens fogged
Rinse lens
Dirty lens.
Withdraw Flexible Intubation Video Endoscope and clean
distal lens with cotton tip applicator and alcohol.
Lens water-damaged.
Send Flexible Intubation Video Endoscope to KARL STORZ
for repair.
Inadequate lighting
Light conducting fibers damaged.
Send Flexible Intubation Video Endoscope to KARL STORZ
for repair.
Light does not function
Not connected to power.
Connect to power.
Complete failure of the unit
No power from power line
Check the electrical supply
Possible lighting up of status display
Power cord connector is not properly connected to power
Push power cord connector firmly into the power cord
cord receptacle
receptacle
System error
Carry out reset (see page 24).
No image, TFT screen is dark
Defective camera electronics
Send Flexible Intubation Video Endoscope to KARL STORZ
for repair.
TFT screen defective
Send TFT screen to KARL STORZ for repair.
Color distortions
White balance correct
Carry out new white balance.
TFT screen defective
Send TFT screen to KARL STORZ for repair.
Incorrect color setting on the monitor
Set the color on the monitor (see page 17).
Image cannot be saved
No memory card inserted
Insert memory card.
Memory card full
Change memory card.
Video stream cannot be played back
No MPEG-4 codec installed
Install an MPEG-4 codec on the PC
on the PC monitor
.