Table 206 /Table 6
Recommended separation distances between portable and mobile RF communications equipment
and the
KARL STORZ Flexible Intubation Video Endoscope 11302 BDX
in conjunction with monitor 8403 ZX
The Flexible Intubation Video Endoscope 11302 BDX in conjunction with monitor 8403 ZX is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the unit as recommended below,
according to the maximum output power of the communications equipment.
Separation distance d [m]
according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Rated maximum
output power of
d=[3.5 ⁄3]
兹苶
P
d=[3.5 ⁄3]
兹苶
P
d=[7 ⁄3]
兹苶
P
transmitter
[W]
0.01
0.1
0.1
0.2
0.1
0.4
0.4
0.7
1
1.2
1.2
2.3
10
3.7
3.7
7.4
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note:
2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Electromagnetic Compatibility (EMC) Information