Warning:
The pressure compensation cap (P/N Z11457)
2&
must be removed from the vent port
e
for:
• endoscopic interventions
• immersing in liquid, e.g. disinfectant solution.
Warning:
Danger of glare and burns! Never look directly into the light of the Flexible Intubation Video Endoscope.
The LED light radiated through the Flexible Intubation Video Endoscope may cause glare or high temperatures at the
distal light outlet and therefore lead to burns.
Warning:
Danger of explosion if flammable gases are used in the immediate proximity of the Flexible Intubation
Video Endoscope. The Flexible Intubation Video Endoscope must not be used in the presence of flammable or
combustible media.
Warning:
The Flexible Intubation Video Endoscope may only be repaired by personnel authorized by KARL STORZ.
Unauthorized repairs may impair functioning and safety, and are therefore strictly forbidden.KARL STORZ gives no
warranty for Flexible Intubation Video Endoscopes that are repaired by unauthorized personnel.
Warning:
Damage to the Flexible Intubation Video Endoscope and injury to the patient or user resulting from
incorrect operation are not covered by the manufacturer’s warranty.
Warning:
Always unplug the Flexible Intubation Video Endoscope before carrying out any maintenance and
cleaning work on it.
Warning:
The Flexible Intubation Video Endoscope must not be used during the discharge of a cardiac defibrillator.
Remove the Flexible Intubation Video Endoscope from the operation area before discharge takes place.
Warning:
When in operation, the integrated LED light source heats up the Flexible Intubation Video Endoscope.
Switch off the Intubation Videoscope before putting it down and do not place it in the proximity of the patient.
Warning:
Only KARL STORZ devices (monitor 8403 ZX) may be used. Disregarding this warning can endanger the
patient and user.
Warning:
If the Flexible Intubation Video Endoscope should malfunction during application on a patient, stop the
application immediately. Place the distal tip of the Flexible Intubation Video Endoscope in the neutral, nondeflected
position and remove the Flexible Intubation Video Endoscope slowly and carefully from the patient.
CAUTION:
Only use accessories (wires, etc.) from KARL STORZ. Accessories from other manufacturers may
damage the Flexible Intubation Video Endoscope.
CAUTION:
Do not steam sterilize (autoclave) the Flexible Intubation Video Endoscopes under any conditions.
CAUTION:
The maximum permissible temperature for decontamination and sterilization is 65° C.
CAUTION:
The pressure compensation cap (P/N Z11457)
2&
must be in place on the vent port
e
for:
Safety Instructions
11