Inspection of applied part:
•
Visual inspection
•
Sensor test
•
Heating resistance test
Inspection of control unit:
•
Visual inspection
•
Test of the cable breakage shut-off
•
Temperature control test (this test is used to check the essential performance
feature)
•
Test of the overheating and low temperature alarm
•
Test of the heating resistance monitoring
•
Electrical safety inspection (in accordance with IEC/EN 62353)
- Protective Earth resistance
- Insulation resistance
- Alternative device leakage current
- Alternative patient leakage current
The tests must be performed according to the specifications in the test
instructions (Scope of delivery of the ASTOPAD test box)
13.2 Replacing the battery
1. Disconnect the device completely from the mains power supply (by discon-
necting the mains plug)
2. Press the "Standby" button until the "Standby" LED switches off.
3. Disconnect the mains plug from the device.
4. Remove the housing cover by loosening the four housing screws (Fig. 6).
5. Remove the battery and push the new battery as far as possible
(Figure 7).
6. Mount the housing cover and secure with the four housing screws (Fig. 6).
Fig. 6
Fig. 7
Battery