12.2 Overview of alarms
Alarm
Display
“Start”
LED
Output
A or B
“Alarm”
LED
“Standby”
LED
Acous-
tic
alarm
signal
Possible
reasons
Green
Yellow
Blue
Low tempera-
ture alarm
“A1”
Actual temperature
Set temperature
A1
Actual temperature
≤1.0 °C set
temperature
Overheating
alarm “A2”
Actual temperature
Set temperature
A2
Actual temperature
≥1.0 °C set
temperature
Time alarm
“A3”
Actual temperature
Set temperature
A3
Set temperature
not reached during
60 minutes of
uninterrupted
heating
Overheating
alarm “A4”
Actual temperature
Set temperature
A4
Actual temperature
≥ 41.0 °C
H
Set temperature
A4
Temperature of the
applied part
> 45.0 °C
Sensor defect
alarm “A5”
Actual temperature
Set temperature
A5
Connection
interrupted
between control
unit and applied
part or sensor
defect in applied
part
Heater defect
alarm “A6”
Actual temperature
Set temperature
A6
Connection
interrupted
between control
unit and applied
part or heater
defect in applied
part
=
LED off
=
LED on
=
LED flashes