1.
Perform the daily cleaning procedure. See
2.
Put a disinfectant solution of 5% non-foaming Sodium Hypochlorite into a shallow
heat-resistant container (this prevents it from being drained at start of cycle).
Disinfection solution
Volume required for tank
5% of non-foaming
Sodium Hypochlorite
5% is equivalent to
2150 ml (72.7 US fl.oz)
3.
Put the container of disinfectant solution in the basket.
4.
Push a luer hose with attached distal connector
into a flush port in all corners of the basket. This
aids circulation to ensure all basket manifolds
are disinfected.
5.
Put the basket in the tank. Make sure the
basket is fully docked and both side catches
engage over the top of the basket.
6.
Close the lid and pull the lid handle forward to
engage lid locks.
86
MAINTENANCE
InnoWave PCF Sonic Irrigator
Operator Manual
Summary of Contents for InnoWave PCF Sonic Irrigator
Page 1: ...InnoWave PCF Sonic Irrigator Operator Manual ENGLISH ...
Page 2: ......
Page 20: ...12 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 86: ...78 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 98: ...90 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 110: ...102 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 120: ...112 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 138: ...130 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 156: ...148 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 160: ...152 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 162: ...154 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 164: ...156 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...