9.
The instrument is now loaded. Repeat the
procedure for instrument loading for up to 10
Endowrist Si/S instruments.
10. It is recommended to have at least one
instrument connected to each flush manifold in
four corners of the basket. If processing less
than 4 instruments, attach a luer hose to basket
manifolds not being used: this aids water
circulation. Put the luer hose through basket
support to prevent it from moving around during
the wash cycle.
5.6.3
Si Multi Basket — Loading Staplers
1.
Inspect the basket. Make sure all flush ports and manifolds are clean and in a good
condition. See
2.
The instrument tip should be open to allow for cleaning of all surfaces.
3.
Each stapler instrument requires connection to
two hoses; a stapler flush adapter(
A
) and a luer
hose (
B
). Make sure both hoses are in good
condition and not ruptured.
52
USING INSTRUMENT BASKETS
InnoWave PCF Sonic Irrigator
Operator Manual
Summary of Contents for InnoWave PCF Sonic Irrigator
Page 1: ...InnoWave PCF Sonic Irrigator Operator Manual ENGLISH ...
Page 2: ......
Page 20: ...12 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 86: ...78 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 98: ...90 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 110: ...102 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 120: ...112 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 138: ...130 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 156: ...148 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 160: ...152 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 162: ...154 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 164: ...156 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...