Cannulated (lumened) instruments should be attached to basket flush ports to
receive flushing during the wash cycle. A standard basket luer hose and a suitable-
sized distal connector will be required to attach each instrument to a basket flush
port.
1.
Inspect the basket. Make sure all four basket
manifolds, located in the corners of the basket,
are clean and in good condition. See
2.
Select a distal connector that best fits over the
end of the instrument. The distal pack, included
with the equipment when shipped, has 4 sizes
of distal connector. The instrument should be
able to push through the silicone opening of the
distal connector and not fall out, but not be held
tightly as the area being clamped will not be
cleaned. Do not push instrument into a distal
connector that is much smaller than the
instrument end, as this could damage the distal
connector silicone opening.
3.
Inspect the distal connector. Do not use the
distal connector if cracked, or if the silicone
opening is damaged.
USING INSTRUMENT BASKETS
39
InnoWave PCF Sonic Irrigator
Operator Manual
Summary of Contents for InnoWave PCF Sonic Irrigator
Page 1: ...InnoWave PCF Sonic Irrigator Operator Manual ENGLISH ...
Page 2: ......
Page 20: ...12 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 86: ...78 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 98: ...90 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 110: ...102 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 120: ...112 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 138: ...130 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 156: ...148 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 160: ...152 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 162: ...154 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 164: ...156 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...