- 23 -
4. INSTALLATION
Your metrology system was carefully secured in a custom crate for stability and protection during
shipment. Exercise care in handling the shipping crate, as excessive force or shock may damage
its delicate contents.
4.1 Required
Tools
The following items are typically required to uncrate and install the metrology system:
1.
Phillips screwdriver
2.
Battery powered drill (Phillips bit recommended)
3.
Standard slotted screwdriver
4.
Small slotted screwdriver for cable connections
5.
Industrial level (2 recommended)
6.
Hex wrenches: metric set, 6 mm and smaller
7.
Needle nose pliers (6” long)
8.
6” crescent wrench (2 recommended)
9.
Black cable ties
4.2 Uncrating
Remove the screws on the top of the crate and carefully remove the top. Remove one of the side
panels for access to the contents. Remove and unpack all components. Verify that the contents
match the packing checklist included in the documentation packet. Once the presence of all items
has been verified, follow the assembly procedures below.
4.3 Placement of AV350 / AV450 Pedestal
The pedestal must be leveled before the metrology unit is placed on top. Leave sufficient clear-
ance in the back and on the sides to access the back of the workstation and metrology unit. A
minimum clearance of 30 cm (12”) is recommended on both sides and in the back for installation
and maintenance. This is in addition to the 30 cm required for travel of the cable track in back of
the metrology unit, and the 25 cm (10”) to the right or left side for the shelf that will hold the
joystick / trackball unit.
4.4 Leveling
of
AV350 / AV450 Pedestal
The pedestal includes a swiveling wheel on all four cor-
ners, plus an adjustable bolt which locks the pedestal in
place and provides limited height adjustment. To lock the
pedestal in place and adjust height, first turn the lock-nut
on the bolt counterclockwise to its upper position. Then
use a wrench to rotate the bolt clockwise to achieve the
desired corner height. When done, rotate the lock-nut
clockwise to prevent the bolt from further rotation. Once
the pedestal has been leveled, carefully lift the metro-
logy unit onto it.
Pedestal leveling mechanism