- 7 -
Feature
AV300-M3
AV350-M3
AV450-M3
Physical Configuration
Benchtop system with
optional workstation for
metrology unit and PC.
Floor standing with machine pedestal and
computer cart for monitor and PC.
X-Y-Z Stage Travel
300 x 150 x 135 mm
(12” x 6” x 5.5”)
350 x 350 x 200 mm
(14” x 14” x 8”)
450 x 350 x 200 mm
(18” x 14” x 8”)
Window Aperture
340 x 198 mm (13.4” x
7.8”)
389 x 389 mm
(15.3” x 15.3”)
420 x 540 mm
(16.5” x 21.3”)
Stage Dimensions
513 x 275 mm (20.2” x
10.8”)
716 x 549 mm (28.2” x 21.6”)
Stage Load Capacity
11 kg (25 lbs)
11 kg (25 lbs)
Base
Cast aluminum
Granite
X-Y Accuracy (E2)
1.9 µm + 5L/1000
2.5 µm + 5L/1000
Z accuracy (E1)
2.5 µm + 5L/1000
2.5 µm + 5L/1000
Scale Resolution
0.1
μ
m
Control System Software
MetLogix 2D or 3D CNC software (as ordered).
Lighting Sources
(as ordered)
1.
Fiber-optic light delivery from 150-Watt quartz halogen lamps to a fixed
ring light, sub-stage light, and optional through-the-lens coaxial light.
2.
LED ring light can be fixed, or Contrast Master quad ring light, LED back
light, and optional through-the-lens coaxial light.
LED brightfield quad light with individually adjustable side illumination.
Camera Types
1.3 or 2 MPixel color CCD (as ordered), USB 2.0 digital interface
Camera Optics
6.5:1 zoom lens,
motorized
12:1 zoom lens, motorized
Computer
24” touch-screen
monitor and PC
Rack-mount PC with 24” touch-screen monitor
Screen Resolution
1920 x 1080 pixels
Computer Controls
Touch-screen, wireless keyboard, wireless mouse
Manual Positioning
Pendant with joystick and trackball.
Zoom Control
Motorized under M3 software control.
Lighting Control
Programmable under M3 software control.
Calibration
To NIST traceable calibration standards.
Optional Items
Quad ring light, auxiliary lenses (0.5X, 1.5X, 2X), parts fixturing, NIST
traceable calibration standards, workstation for AV300 system.
Quad Ring Light Option
Lighting Elements
Ring Dimensions
Height Adjustment
96 white LEDs in 4 individually controllable quadrants.
125 mm (5”) OD, 100 mm (3.9”) ID, 19 mm (0.75”) H.
Motorized on Z-track with CNC control by M3 software. Height adjustable
from bottom of video probe to 100 mm (4”) below.