10
· ErgoOne · User Manual
5. Technical Data
Technical Data ErgoOne
®
Pipettes
ErgoOne
®
Single-Channel
Pipette
volume
Systematic measure-
ment deviation
(inaccuracy) ± %
Coincidental measure-
ment deviation
(imprecision; CV) ± %
EN ISO
8655-2
ErgoOne
®
EN ISO
8655-2
ErgoOne
®
0.1 – 2.5µl
0.1µl
1.25µl
2.5µl
±125 %
±10 %
±5.0 %
±24%
±2.5%
±1.4%
±75 %
±6.0 %
±3.0 %
±12%
±1.5%
±0.7%
0.5 – 10µl
1.0µl
5µl
10µl
±12 %
±2.4 %
±1.2 %
±2.5%
±1.5%
±1.0%
±8.0 %
±1.6 %
±0.8 %
±1.8%
±0.8%
±0.4%
2 – 20µl (micro cone)
2µl
10µl
20µl
±10 %
±2.0 %
±1.0 %
±5.0%
±1.2%
±1.0%
±5,0 %
±1,0 %
±0,5 %
±1.5%
±0.6%
±0.3%
2 – 20µl (standard cone)
2µl
10µl
20µl
±10 %
±2.0 %
±1.0 %
±5.0%
±1.2%
±1.0%
±5,0 %
±1,0 %
±0,5 %
±1.5%
±0.6%
±0.3%
10 – 100µl
10µl
50µl
100µl
±8.0%
±1.6 %
±0.8 %
±3.0%
±1.0%
±0.8%
±3.0 %
±0.6 %
±0.3 %
±1.0%
±0.3%
±0.2%
20 – 200µl
20µl
100µl
200µl
±8.0 %
±1.6 %
±0.8 %
±2.5%
±1.0%
±0.6%
±3.0 %
±0.6 %
±0.3 %
±0.7%
±0.3%
±0.2%
100 – 1000µl
100µl
500µl
1000µl
±8.0 %
±1.6 %
±0.8 %
±3.0%
±1.0%
±0.6%
±3.0 %
±0.6 %
±0.3 %
±0.6%
±0.2%
±0.2%
500 – 5000µl
500µl
2500µl
5000µl
±8.0 %
±1.6 %
±0.8 %
±2.4%
±1.2%
±0.6%
±3.0 %
±0.6 %
±0.3 %
±0.6%
±0.25%
±0.15%
ErgoOne
®
Multi-Channel
Pipette
volume
Systematic measure-
ment deviation
(inaccuracy) ± %
Coincidental measure-
ment deviation
(imprecision; CV) ± %
EN ISO
8655-2
ErgoOne
®
EN ISO
8655-2
ErgoOne
®
0.5 – 10µl
1.0µl
5µl
10µl
±24 %
±4.8 %
±2.4 %
±8.0%
±4.0%
±2.0%
±16 %
±3.2 %
±1.6 %
±5.0%
±2.0%
±1.0%
10 – 100µl
10µl
50µl
100µl
± 16 %
± 3.2 %
± 1.6 %
±3.0%
±1.0%
±0.8%
± 6.0 %
± 1.2 %
± 0.6 %
±2.0%
±0.8%
±0.3%
30 – 300µl
30µl
150µl
300µl
±26.7 %
±5.3 %
±2.7 %
±3.0%
±1.0%
±0.6%
±10 %
±2.0 %
±1.0 %
±1.0%
±0.5%
±0.3%
Measurement data determined in compliance with the test conditions for piston-stroke pipettes
according to EN ISO 8655
Measurement carried out with original TipOne
®
pipette tips