Volume range ErgoOne
®
ErgoOne
®
model
Volume range (min./ max.)
Colour code
2.5µl
0.1 – 2.5µl
Red
10µl
0.5 – 10µl
Red
20µl (micro cone)
2 – 20µl
Red
20µl (standard cone)
2 – 20µl
Yellow
100µl
10 – 100µl
Yellow
200µl
20 – 200µl
Yellow
300µl
30 – 300µl
Green
1000µl
100 – 1000µl
Blue
5000µl
500 – 5000µl
Violet
10µl Pipette
10.00µl
0.50µl
20.0µl
100µl
200.0µl
1000µl
200µl Pipette
1000µl Pipette
Examples of typical volume displays:
4
· ErgoOne · User Manual
2.2 Setting the Volume
Open the volume adjustment catch by
pulling up the volume control button
1
(Fig. 1), until you hear a click. Turn
the volume control button
1
clockwise
to reduce the aspirated volume, or
counter clockwise to increase it. The
set volume is shown in the four-step
volume display
4
from top to bottom,
with a white line indicating the decimal
point. When you have finished setting
the volume, push the volume control
button
1
back to its lock position until
you hear a click. The pipette is now
ready to use.
Please note the minimum
and maximum volume
range of your pipette (Ta-
ble 1). You have reached
the end point of the vol-
ume adjustment when it
becomes more difficult to
turn the button and you
can hear a clicking sound.
Do not use excessive force!
Table 1
Fig. 1