Pubn No: 30000H0K
10. SPECIFICATION
Flash point temperature range: Ambient to 300°C
0°C to 300°C using optional Seta Coolant
Module
Sample size: 2ml for flash points up to 100°C
4ml for flash points above 100°C
2ml for FAME products at all temperatures.
Test time: 1 minute for flash points up to 100°C
2 minutes for flash points above 100°C
1 minute for FAME products at all temperatures
Timing: Integral timer for standard 1 and 2 minute tests,
and up to 30 minutes
Audible and visual indication
Temperature measurement:
(calibrated platinum resistance thermometer)
0.5°C resolution
0.5°C accuracy
Flash detection: Automatic
Sample Cup material Aluminium
Operational ambient temperature range: 5 - 40°C
Supply voltage: 115 or 220V (
!
15%) (switchable) 50/60Hz
(12Vdc with optional 30005-0 PSU)
Power consumption: 200W
Supply fuse: 2A antisurge (T) HBC type
Dimensions (H x W x D): 195 x 295 x 140mm
Weight: 3kg
Issue K
10 Jun 2007
34