Pubn No: 30000H0K
Wait until the temperature shown
on the thermometer is stable.
This may take a few more
minutes.
54.9
°C
55.1
°C
If the thermometer shows 54.9 to
55.1°C the calibration is correct.
Otherwise repeat the calibration
procedure.
Switch OFF
Switch the instrument OFF at the
ON/OFF Power Switch on the
rear of the instrument. The
calibration is stored.
Remove and clean the
Thermometer.
Issue K
10 Jun 2007