68
To change batteries proceed as follows:
1.
Remove the screws at the back of the remote control.
2.
Open the case and remove the plastic holder and remove the batteries.
CAUTION: Be careful with the circuit board.
3.
Replace the batteries with new ones. Make sure the polarity is correct. Polarity markings are
located at the bottom of the casing.
CAUTION: Do not reverse the poles.
Replace the plastic holder and close the case. Secure it with the screw.
System Backup Battery
The internal system backup battery is recharged by leaving the QUANTIEN™ measurement
system turned on. The battery is not intended to be replaced.
Disposal
NOTE: To avoid environmental damage and/or personal injury, handling and disposal of
batteries and electronic equipment shall be done in accordance with applicable local,
state and federal laws and regulations.
Cleaning
CAUTION: Do not immerse in liquid: Do not use the QUANTIEN™ measurement
system if it has been immersed in liquid.
1.
Turn OFF the unit and disconnect the mains power cable.
2.
After each patient, clean the exterior with a moist tissue, using regular cleaning agents, or if
necessary, alcohol, and wipe dry.
3.
Reconnect the main power cable.
4.
Press ON/OFF switch to ON.
WARNING:
Make sure that every part of the unit is dry before reconnecting to the power
supply.
Do not use the QUANTIEN measurement system if liquid is suspected to
have penetrated the enclosure or the power supply unit. This can cause the
user or patient to be exposed to electric shock. Contact St. Jude Medical
representatives or technical support for further instructions.
Preventative Maintenance Schedule
It is recommended that the scheduled preventative maintenance of the QUANTIEN™
measurement system should be performed with an interval not exceeding 12 months. The list
below shows the operations needed to ensure the electrical safety and measurement accuracy of
the QUANTIEN measurement system. These operations are preferably performed in the order
according to the list below. Instructions for performing these operations are found in the
instructions in the remainder of this chapter.
The measurements should always be carried out by a qualified technician.
Ageing of cabling and multi-use pressure transducers can affect performance of measurements. It