63
Table 30. ECG signal on QUANTIEN™ measurement system
Erroneous
function
Possible cause
Check
Normal function Corrective action
Wrong type of
AUX IN cable
AUX IN cable
compatibility with
the present HRS
HRS listed on the
actual AUX IN
cable
Replace with a
correct AUX IN
cable
Table 31. Pressure signal on the HRS via PressureWire™ OUT
Erroneous
function
Possible cause
Check
Normal function Corrective action
Unable to zero, or
abnormal or
absent PW OUT
pressure signal
on the HRS
PW OUT not
properly
connected to
HRS input
Connection of
monitor cable
Properly
connected
Correct the failing
connection
Bad contact in
cable connectors
Stress the actual
connection
mechanically
The pressure
signal on the HRS
will remain
correct
Replace the
broken cable
Wrong type of
monitor cable
Monitor cable
compatibility with
the present HRS
HRS listed on the
actual monitor
cable
Replace with a
correct monitor
cable
Defective monitor
cable
Visually check the
monitor cable
Strain relief intact,
no other damage
Replace the
monitor cable
Defective PW
OUT
Test PW OUT
according to
Testing PW Out
(page 71)
Test of PW OUT
passed without
failure
Contact St. Jude
Medical
Defective HRS
input channel
Test PW OUT
according to
Testing PW Out
(page 71)
Test of PW OUT
passed without
failure
Contact
manufacturer of
the HRS
PW OUT not
configured
correctly
Configuration of
PW OUT in
System Menu
PW OUT
configured
correctly
Edit and correct
PW OUT
configuration in
System Menu
Table 32. Pressure signal on the HRS via AUX OUT
Erroneous
function
Possible cause
Check
Normal function Corrective action
Abnormal or
absent pressure
signal on the HRS
AUX OUT not
properly
connected to
HRS input
Connection of
AUX cable
Properly
connected
Correct the failing
connection
State: Released