When demo mode is activated a pre-recorded sequence of pressure tracings including a
hyperemic event is replayed on the screen allowing you to perform the steps of an FFR
measurement procedure.
Any system setup changes made in demo mode will be discarded when the QUANTIEN™
measurement system is turned off or if you are exiting demo mode to normal mode. The settings
can be saved in the
Service
menu, select
Save setup changes
.
Troubleshooting
On Screen Messages
The tables below list on screen messages. Messages that are self-explanatory and information of
minor importance are not included.
Letter in first column indicates type of message.
E = Error
W= Warning
I= Information
Message tables:
System on-screen messages (page 47)
AO Source — Wi-Box™ AO transmitter (page 50)
AO Source — AUX IN (page 51)
PW Source — PressureWire™ Wireless guidewire (page 51)
PW Source — PressureWire™ Cable-Connection guidewire (page 52)
Table 19. System on-screen messages
Typ
e
Current Message
Reason
Action
Cancel
OK
E
"Failed to connect to
DICOM peer"
Request (or
query) for
DICOM worklist
failed, i.e., no
response from
remote host
Check
settings
Check
network
connection
Check
remote host
N/A
Acknowledges
message and
closes window
I
"There were more patients
matching the worklist
query available. Use a
more specific query"
Query to remote
host contains
more records
than will be
displayed
Redefine
query
N/A
Acknowledges
message and
closes window
W
"Patient <Last, First> has
studies associated that are
write protected, they will
not be updated"
When an edited
patient has
write protected
studies
Select
Cancel or OK
Discards
edits and
closes
window
Confirms edits
and closes
window
State: Released
Date: 2019.03.06 22:30 GMT