Table 34. Export function
Erroneous
function
Possible cause
Check
Normal function Corrective action
Error in network
configuration
Network
configuration in
System menu
N/A
Correct network
configuration in
System menu
DICOM export
configured
incorrectly
Run verify
function in
DICOM Export
setup in System
menu
Verify successful Correct DICOM
Export
configuration
No export at end
of recording
Auto export not
configured
Auto export
configuration in
system menu
(Export)
N/A
Correct auto
export
configuration
Not possible to
open exported
files in RadiView
software
Exported file
format not
compatible with
all versions of
RadiView software
Compatibility
mode setting in
System Menu
(Export)
N/A
Set flag for export
in compatibility
mode
Table 35. External display function
Erroneous
function
Possible cause
Check
Normal function Corrective action
Disturbances or
no QUANTIEN™
display viewed on
external monitor
Video OUT (DVI-I)
not properly
connected to
external monitor
input or bad
contact in cable
connectors
Connection of DVI
cable. Stress the
actual connection
mechanically
QUANTIEN
display viewed on
external monitor
Correct the failing
connection or
replace broken
cable
Wrong cable
Compatibility of
cables and
adapters
according to
Specifications
Interfaces (page
75)
N/A
Replace with a
correct cable
and/or adapter
Defective cable
Visually check the
monitor cable
Strain relief intact,
no other damage
Replace the cable
Defective DVI-I
port
Test with another
external monitor
QUANTIEN
display viewed on
external monitor
Contact St. Jude
Medical
State: Released