17
4.
Press the
Accept
button to save room settings.
For detailed information on Room settings, see Room Setup (page 32).
Regional Settings
To set language, time zone, date and time go to
(Setup)
Menu, press
System
, press
Regional
. See Regional Settings (page 17) for details.
Measure FFR or RFR
Warnings and Precautions
WARNING:
Do not use the QUANTIEN™ measurement system if it has been dropped, or
in any other way exposed to mechanical or electrical damage, or if liquids are
suspected to have penetrated the casing or the power supply unit. This can
cause the user or patient to be exposed to electric shock or cause faulty
readings of data. Contact St. Jude Medical representative or technical
support for further instructions.
Do not use the QUANTIEN measurement system if there is reason to believe
the system's security has been compromised or if the system was
unaccounted for a period of time (i.e. misappropriated, modified or tampered
with).
The operator should not touch QUANTIEN measurement system non-CF
connectors (or other non-medical equipment) and the patient or patient
leads at the same time. Conductive connection may cause leakage currents
to induce ventricular fibrillation.
CAUTION: The QUANTIEN measurement system is Type CF equipment and
protected against the effects of a discharge of a defibrillator. PressureWire™
guidewire readings may be affected by defibrillation. Recalibrate PressureWire
guidewire after defibrillation. After defibrillation restart procedure, i.e. re-zero and
re-equalize PressureWire guidewire.
State: Released
Date: 2019.03.06 22:30 GMT