48
Table 19. System on-screen messages
Typ
e
Current Message
Reason
Action
Cancel
OK
W
"Patient <Last, First> has
several studies associated,
mark the ones you wish to
delete in the list below"
When trying to
delete a patient
who has more
than one study
in the archive
Select
Cancel or OK
No delete
and close
window
Confirm delete
and close
window
I
"Export failed"
Export failed
Check
settings for
export jobs.
Check
network
connection
Check USB
N/A
Confirm
W
"Register new patient?"
Changing the
PressureWire™
guidewire for an
active patient
Select
Cancel, Yes
or No
Continue
with the
current
PressureWire
guidewire.
Close
window
Yes: Switches
to Patient
Screen
No: Changes
the
PressureWire
guidewire but
keeps the
current
patient
W
"A patient is registered?
Do you wish to proceed
anyway?"
Entering demo
mode with an
active patient
Select
Cancel or OK
Cancel
operation
and close
window
Continue to
enter demo
mode, close
window
E
"Failed to load recording
header. File is missing or
corrupt."
Header file is
missing or
corrupt, e.g.
with checksum
error.
Contact
St. Jude
Medical
N/A
Acknowledges
message and
closes window
E
"Failed to load recording
data. File might be
corrupt."
Data file is
missing or
corrupt, e.g.
with checksum
error.
Contact
St. Jude
Medical
N/A
Acknowledges
message and
closes window
E
"Failed to load recording
data. Unable to locate
data file."
Data file is
missing or
corrupt, e.g.
with checksum
error.
Contact
St. Jude
Medical
N/A
Acknowledges
message and
closes window
State: Released