UGA Series Universal Gas Analyzers
Remote Programming
3 –
ZMBK (?) {i}
ZMOD?
System Bake mode
The ZMBK i command turns the system bake mode Off
(i=0) or On (i=1). The ZMBK? command returns whether
the system bake mode is On or Off.
Example
ZMBK
starts the system bake.
ZMBK?
returns the state of the Bake mode, On(1) or Off(0).
Errors and warnings
The command is ignored if any of these conditions are true.
Warning
47
“Leak Test ON”
Warning
51
“Pressure Check off”
System mode query
The ZMOD? Query returns the current UGA mode
described below.
Off
(1)
all components are OFF
START
(3)
Automatic START sequence in prog-
ress
STOP
(4)
Automatic STOP sequence in prog-
ress
READY
(6)
RGA is on
SLEEP
(7)
Automatic SLEEP sequence in prog-
ress
IDLE
(8)
TP and RP are idle, all other compo-
nents are off.
LEAK TEST
(0) Leak test mode is ON
SYSTEM BAKE
(2) System Bake is ON
MANUAL
(3) Manually changed of the state of a
component from a pre-defined state