UGA Series Universal Gas Analyzers
Guide to Operation
2 – 1
2.3.3 Connection to a PC
2.3.3.1 TCP/IP Ethernet Connection
In order to use this program, at least one UGA must be connected
to the PC. There are two ways to connect the UGA to a PC; TCP/IP
Ethernet or an RS232 serial connection. In this section, we provide
step-by-step instructions for setting up communication.
In order to use this connection, the user should know the network
environment for the UGA and the PC. The UGA firmware and
software both need proper TCP/IP parameters.
.
First, determine the IP address, Subnet Mask, and Gateway
for communication. The following are default values.
Refer to your network administrator for appropriate
values for your network.
IP address : 0.0.0.0
Subnet Mask : 255.255.255.0
Gateway :
0.0.0.0
2.
Once you obtain these parameters, enter them through the
front panel with the following procedure. Alternatively, a
user may type in these parameters through RS232 serial
communication. (Refer to the previous section)
From the front panel;
•
Power up the UGA.
•
Push the “Level-up” button to enter MENU at the front
panel.
•
Select “Communication” and hit the confirmation button.
•
Ethernet items are now shown on the display.
•
Choose “IP Address”. The display shows the default IP
address.
•
Using the “Up” or “Down” button, change the value Then
hit the confirmation button,
.
•
The next number will be blinking. Repeat as above to set
the proper value.
•
After keying in all four numbers, the display will move
up to the Ethernet menu. Choose “Subnet Mask” and set
the parameters.