94
Remote Programming
ERROR STATUS BYTE
bit
name
usage
0
Plot Error
Set when an error occurs during plotting
1
Print Error
Set when an error occurs during a print screen
operation
2
Memory Error
Set when a memory error is detected
3
Disk Error
Set when an error occurs during a disk or file
operation
4
Unused
5
Clock Unlock
Set when the internal bin clock time base is not
valid
6
Rate Error
Set when a trigger arrives while the previous
record is still being acquired or accumulated
7
Overflow
Set when the count in any bin overflows during
acquisition or accumulation
The Error Status bits stay set until cleared by reading or by the *CLS command.