31
Mode Menu
Ext
1
Yes No
ReAcquire:
Add Toggle
0
Return:
Accumulate
Pause at Toggle
From Offset
0.00
µ
sec
Toggle Cnt.
Accumulate Mode
The Accumulate key selects the accumulation mode. ADD mode sums
new record data to the accumulation. The maximum accumulation per
bin is 32,767. EXT and TOGGLE can either add to or subtract from the
accumulation. When either of these modes is selected, the maximum ac-
cumulation per bin is ±16,383.
EXTernal mode means that the SUBTRACT input will determine whether
each record is added or subtracted. If SUBTRACT is high when a record
is triggered, then that record's data is subtracted from the accumulation.
TOGGLE mode will switch between ADD and SUBTRACT after every
Toggle Count number of records. In this mode, the TOGGLE output will
be TTL high when the next record is to be subtracted and TTL low when
the next record is to be added. The TOGGLE output switches at the end
of SYNC/BUSY to reflect the accumulate mode of the next triggered
record. When data acquisition is started, the first toggle count number of
records is added before switching to subtract. The Add/Sub status indica-
tor on the screen shows the toggle status during a scan. Toggle is gener-
ally used for background subtraction.
The Trigger Offset entry field accepts numeric entry or knob adjustments.
The resolution is 16 bins and the limit is 16,320 bins. Below the entry
field, the actual time corresponding to the trigger offset is displayed. This
is simply the Trigger Offset (in bins) multiplied by the bin width.
The Trigger Offset is used when the data of interest occurs long after the
trigger and short bin widths are required for resolution.
More
The More key displays the second Mode menu. This menu sets the ac-
cumulation mode and also reacquires displayed data from the data ac-
quisition memory.
Summary of Contents for SR430
Page 2: ......
Page 6: ...4 ...
Page 22: ...20 Guide To Operation ...
Page 26: ...24 Guide To Operation ...
Page 36: ...34 Mode Menu ...
Page 54: ...52 Save Menu ...
Page 60: ...58 Recall Menu ...
Page 70: ...68 Plot Menu ...
Page 74: ...72 Test Menu ...
Page 76: ...74 Info Menu ...
Page 97: ...96 Remote Programming ...
Page 99: ...98 98 Program Examples ...
Page 107: ...106 106 Test and Calibration ...
Page 113: ...112 112 Using Photomultiplier Tubes ...
Page 125: ...124 124 124 Circuit Description ...