60
Math Menu
Pressing a smoothing key will smooth the data with the corresponding
width. A width of 25 points means that features which span less than 25
bins will be smoothed out. The time displayed in each menu box is the
smoothing width multiplied by the bin width and represents the real time
smoothing width.
All of the data record in the display memory is smoothed, regardless of
the graph display. While calculations are in progress, the "Calculating"
message is displayed below the graph. Smoothing may be repeated until
the desired noise reduction is attained.
Return
The Return key will return to the main Math menu.
25 pt.
21 pt.
17 pt.
11 pt.
5 pt.
25 pt:
256
µ
sec
21 pt:
215
µ
sec
17 pt:
174
µ
sec
11 pt:
113
µ
sec
5 pt:
51.2
µ
sec
Return:
menu. Once smoothed, the data remains in floating point form even
when the Smooth menu is exited. If floating point data is saved to disk
using the Save Trace function, it will be recalled as floating point data us-
ing Recall Trace. To avoid accidentally losing original data, save the
count data to a disk file before performing Smooth menu opera-
tions.
Return
MATH
Smooth
25 pt
21 pt
17 pt
11 pt
5 pt
Summary of Contents for SR430
Page 2: ......
Page 6: ...4 ...
Page 22: ...20 Guide To Operation ...
Page 26: ...24 Guide To Operation ...
Page 36: ...34 Mode Menu ...
Page 54: ...52 Save Menu ...
Page 60: ...58 Recall Menu ...
Page 70: ...68 Plot Menu ...
Page 74: ...72 Test Menu ...
Page 76: ...74 Info Menu ...
Page 97: ...96 Remote Programming ...
Page 99: ...98 98 Program Examples ...
Page 107: ...106 106 Test and Calibration ...
Page 113: ...112 112 Using Photomultiplier Tubes ...
Page 125: ...124 124 124 Circuit Description ...