If all standards are measured, you get the coefficient of the calibration K and B as well as the coefficient of
determination R.
10.5.4 Use an existing calibration curve
1.
Press [Function]. (Press again, if necessary, to display
Standard Curve
menu item).The LCD will
display as blow.
2.
Press [6] to select
Standard Curve
menu item. Press [Enter]. The LCD will display as blow.
3. Press [2] to select
Load Curve
menu item. Press [Enter].
The LCD will display as blow.
4.
Use digital key to load
curve stored in spectrometer
, press [Enter].
10.5.5 Delete an existing calibration curve
1.
Press [Function]. (Press again, if necessary, to display
Standard Curve
menu item).The LCD will
display as blow.
5
.
Lamp on/off
6
.
Standard Curve
7
.
Language
8
.
Com. With PC
Curve No.:
1
.
Create Curve
2
.
Load Curve
3
.
Delete Curve
4
.
Print Curve
5
.
Lamp on/off
6
.
Standard Curve
8
.
Com. With PC
7
.
Language
2.
Press [6] to select
Standard Curve
menu item. Press [Enter]. The LCD will display as blow.
3. Press [3] to select
Delete Curve
menu item. Press [Enter].
The LCD will display as blow.
Curve No.:
1
.
Create Curve
2
.
Load Curve
3
.
Delete Curve
4
.
Print Curve
4.
Use digital key to delete
curve stored in spectrometer
, press [Enter].
User’s Manual SP-UV 200 23
Spectrum Instruments