4. Remove the two screws (11) and the lamp house cover (12) on the underside of the instrument.
Fig. 4 Lamp compartment of SP-UV 200, opened
5. Unscrew the holder of the halogen lamp (13) and take the holder with the halogen lamp out of the
device.
Fig. 5 SP-UV 200holder of the halogen lamp
6. Remover the lamp (14) from lamp holder
7. Replace the new lamp carefully in the holder.
Caution
Do not handle Halogen lamp with bare fingers. Use a piece of tissue or cloth when handling the lamp. The
oil from your fingers can cause the lamp to burn out prematurely!
8. Replace and fix the holder with the new halogen lamp in the device
9. Close the cover of the lamp house using the 2 screws.
10. Replace the device with the top side on top.
7.2 Replacing Deuterium lamp
Risk of electric shock!
Before changing the deuterium lamp, make sure to switch the device off with the power switch.
Caution: Risk of burns!
Allow the deuterium lamp to cool down to avoid the risk of burns!