5.
Press [6] to select
Standard Curve
menu item. Press [Enter].The LCD will display as blow.
6.Press [1] to select
Create Curve
menu item. Press [Enter].
The LCD will display as blow.
7.Press [1] to select
No. of samples
menu item. Press [Enter].
The LCD will display as blow.
Input
Sample Num:
1
.
No. of Samples
2
.
Unit
3
.
Input Conc
4
.
Test Sample
1
.
Create Curve
2
.
Load Curve
3
.
Delete Curve
4
.
Print Curve
5
.
Lamp on/off
6
.
Standard Curve
7
.
Language
8
.
Com. With PC
8.
Use digital key to enter
samples number
, press [Enter].
9. Press [2] to select
Unit
menu item. Press [Enter].
The LCD will display as blow.
10.
Use digital key to enter
sample unit
, press [Enter].
11. Press [3] to select
Input Conc
menu item. Press [Enter].
The LCD will display as blow.
Input
Conc:
Number:1
1
.
mg/L
2
.
ug/L
3
.
ng/kg
4
.
ug/kg
12.
Use digital key to enter
sample concentration
, press [Enter].
13. Press [4] to select
Test Sample
menu item. Press [Enter].
The LCD will display as blow.
Number: 1
Conc: 1.00
Abs.:
14.Insert the standard solutions in the light beam
. Press [Enter].
User’s Manual SP-UV 200 22
Spectrum Instruments