Atlas™ High Temperature Film Maker Accessory
27
necessary to do so, a small flat screw driver blade can be used in this
groove to prise the spacer ring away from its location if it has become
stuck fast (e.g. from a flow of excess molten sample that has cooled on
the spacer ring and polished surface), from an incorrect film making
procedure. (See
Fig 15.)
Fig 15. Spacer Ring Positioned in Lower Heated Pressing
Surface Assembly
With the Spacer ring (
18
) correctly positioned on the lower polished
surface (
14
), using the forceps supplied place either a flat aluminum
foil disc (
21
) or an already prepared aluminium foil sample cup (
22
)
inside the spacer ring (
18
). When using a sample cup (
22
), make sure
it is centered within the spacer ring (
18
). Where interference fringes
are required the shiny surface of the aluminium foil should be in
contact with the sample to be heated and pressed. (See
Fig 16.
)
Take the sample to be melted and pressed and place it carefully into
the centre of the lower foil disc (
21
) or sample cup (
22
). The amount of
sample to use will vary depending on the thickness of spacer ring (
18
)
chosen. (See
Fig 17.
)
14
18
5