85
--Response message:
Status value, voltage value, current or resistance value
----Program data< status >:
0 (READY)
1 (TEST)
2 (PASS)
3 (FAIL)
4 (STOP)
----Program data< voltage value >:
Data format: Integer
Resolution: 1
Unit: V
----Program data< current or resistance value >:
Data format: float
Unit: mA/MΩ
--Example:
:TEST:FETCH2?
If the current monitor voltage is 1000V, and the current monitor current
is 1.00mA in an ACW test,
1, 1000, 1.00 is returned.
:ALLSET3
Sets the function and the parameters of the first test step.
--Syntax:
1.
Setup ACW parameters
:ALLSET3
<func>,<scan>,<volt>,<upper>,<lower>,<time>,<arc>,<real>,
<freq>,<rise>,<fall>,<offset>
12 items, see parameters list.
2.
Setup DCW parameters
:ALLSET3
<func>,<scan>,<volt>,<upper>,<lower>,<time>,<arc>,<rise>,
<fall>,<wait>,<check>,<offset>
12 items, see parameters list.
3.
Setup IR parameters
:ALLSET3
<func>,<scan>,<volt>,<upper>,<lower>,<time>,<rise>,<fall>,
<sagc>,<range>,<offset>
11 items, see parameters list.
4.
Setup OS parameters