20
Figure 3-3 AC voltage load adjusting rate
DC
withstanding
voltage
test
5kV/10mA(ST9201/S)
5kV/5mA(ST9201B/C)
ST9201 series can provide DC withstanding voltage test of wide voltage range
(Max. output voltage is 6kV). Instrument allocates a reliable, low ripple
DC/AC switch circuit, voltage load adjusting rate≤1%+10V. When the
voltage is between 50-500V, low load impedance may cause unstable output
voltage, set the hardware automatic voltage regulation function as OFF and
the voltage regulation factor is 10%.
Figure 3-4 DC voltage output range