68
Also inquires about the test voltage.
--Syntax:
Command message: :
SOUR:SAFE:STEP <sn>:DC:LEV <volt>
Query message:
:SOUR:SAFE:STEP <sn>:DC:LEV?
--Program data<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit:
--Program data<volt>:
Data format: float
Set value: 50~6000
Resolution: 1
Unit: V
--Example:
To set the test voltage for step 1 to 1000,
:SOUR:SAFE:STEP 1:DC:LEV 1000.
--Example:
:SOUR:SAFE:STEP 1:DC:LEV?
If the current test voltage for step 1 is 1000,
1000 is returned.
:SOURce:SAFEty:STEP:DC:LIMit:LOW
Sets the LOWER current for DCW test.
Also inquires about the LOWER current.
--Syntax:
Command message:
:SOUR:SAFE:STEP <sn>:DC:LIM:LOW <current>
Query message:
:SOUR:SAFE:STEP <sn>:DC:LIM:LOW?
--Program data<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit: