Wave User Manual
13
Version 4
3.4
The UT Connectors
The UT connectors are located on top of the unit. There is one TX/RX connector and one
RX connector. They are identified directly on the casing. These UT connectors can be
LEMO or BNC.
When using the instrument in pulse-echo mode, only the TX/RX output is used. For pitch
and catch, both TX/RX and RX must be hooked up to the probe.
Figure 4
–
Wave connectors and the expansion ports
3.4.1
The UT Connector Pinouts
The following tables describe the pinouts of the UT connectors. Figure 5 and Figure 6
illustrate the localisation of the pinouts in the Lemo connector and BNC.
Table 1
–
TX/RX connector pinouts
Pin number
I/O
I/O type
Description
1
In/Out
Analogue
Ultrasound pulse transmitter/receiver
2
Out
Power
Ground
Table 2
–
RX connector pinouts
Pin number
I/O
I/O type
Description
1
In/Out
Analogue
Ultrasound pulse receiver
2
Out
Power
Ground
Expansion
GPIO
RX
TX/RX