FEATURES
Automatic power off
In order to conserve power, the multimeter automatically switches off
after approx. 15 minutes if no measurement is performed or function is
changed. This is indicated by 5 short beeps 1 minute before switching
off then the unit switches into sleep mode with a beep. Pressing any
button will wake up the clamp meter.
DC voltage measurement (
)
Connect the red probe cable to the “INPUT” socket, and the black one
to the “COM” socket. Set the dial to the DC voltage position. Connect
the measurement probes to the circuit to be measured. Note: The read
-
ing can be unstable, especially in the 400mV measurement range, if
the probe cable is not connected to the circuit being measured. If the
measured figure is more than 600 V, the “OL” symbol will appear on the
LCD display, and the built-in buzzer will emit a beep.
Measurement limit
Resolution
Accuracy
400 mV
0.1 mV
±(1% 2 digits)
4 V
1 mV
±(0.7% 2 digits)
40 V
10 mV
400 V
100 mV
600 V
1 V
±(0.8% 2 digits)
Input impedance: 10 MΩ
Maximum input voltage: 600 V DC
DC measurement (
)
Warning! Before performing a measurement, unplug the probe cables
from the unit. Set the dial to the desired measurement range. Use the
“
SEL
” pushbutton to set the direct current (
DC
) function. Clear the
display by pressing the “
REL
” pushbutton. Open the clamp, place it
around the cable then close it. The measured figure can be read off
the display. Note: Only measure one conductor at a time. For optimum
results, center the conductor in the clamp meter.
Measurement limit
Resolution
Accuracy
40 A
10 mA
±(3% 6 digits)
400 A
100 mA
Max input current: 400A DC
Measurement of AC voltage (
)
Connect the red probe cable to the “INPUT” socket, and the black one
to the “COM” socket. Set the dial to the AC voltage position. Connect
the measurement probes to the circuit to be measured. Note: If the
measured figure is more than 600 V RMS AC, the “OL” symbol will ap
-
pear on the LCD display, and the built-in buzzer will emit a beep.
Measurement limit
Resolution
Accuracy
4 V
1 mV
±(0.8% 3 digits)
40 V
10 mV
400 V
100 mV
600 V
1 V
±(1% 4 digits)
Input impedance: 10 MΩ
Maximum input voltage: 600 V DC or 600 V AC RMS
Measuring range: 40 Hz- 400 Hz
Measurement of AC current (
)
Warning! Before performing a measurement, unplug the probe cables
from the unit. Set the dial to the desired measurement range. Use the
“SEL” pushbutton to set the alternating current (AC) function. If the
displayed figure is unstable, clear the display by pressing the “REL”
pushbutton. Open the clamp, place it around the cable then close it.
The measured figure can be read off the display. Note: Only measure
one conductor at a time. For optimum results, center the conductor in
the clamp meter.
Measurement limit
Resolution
Accuracy
40 A
10 mA
±(3% 6 digits)
400 A
100 mA
Max input current: 400 A AC
Measuring range: 40 – 400 Hz
Frequency measurement (Hz)
In the frequency measurement function:
Connect the red probe cable to the “INPUT” socket, and the black one
to the “COM” socket. Set the dial to the frequency measurement func
-
tion (HZ%). Use the “Hz%” button to set the frequency measurement
function (Hz). Connect the measurement probes to the circuit to be
measured. Note: Do not exceed the input voltage, as this will damage
the instrument.
Measurement limit
Resolution
Accuracy
9.999 Hz
0.001 Hz
±(0.5% 3 digits)
99.99 Hz
0.01 Hz
999.9 Hz
0.1 Hz
9.999 kHz
1 Hz
99.99 kHz
10 Hz
999.9 kHz
0.1 kHz
9.999 MHz
1 kHz
Input voltage: 200 mV - 10V AC RMS
Over-voltage protection: 250 V DC or 250 V AC RMS
When measuring AC:
Warning! Before performing a measurement, unplug the probe cables
from the unit.
Set the dial to the desired measurement range (40
, 400
).
Use the “SEL” pushbutton to set the alternating current (AC) function.
Use the “Hz%” button to set the frequency measurement function (Hz).
Open the clamp, place it around the cable then close it.
Measurement limit
Resolution
Accuracy
10 Hz
0.01 Hz
±(1.5% 5 digits)
1 KHz
1 Hz
>1 KHz
1 Hz
Measurement limit: 10 Hz – 1 KHz
Input current range: ≥ 4 A AC RMS (higher input current is required at
a higher frequency)
Max input current: 400 A AC RMS
When measuring AC voltage:
Connect the red probe cable to the “INPUT” socket, and the black one
to the “COM” socket. Set the dial to the AC voltage position (
).
Use the “Hz%” button to set the frequency measurement function (Hz).
Connect the measurement probes to the circuit to be measured.
Measurement limit
Resolution
Accuracy
10 Hz
0.01 Hz
±(1.5% 5 digits)
1 KHz
1 Hz
10 KHz
1 Hz
>10 KHz
1 Hz
The accuracy of the
measurement result is
not guaranteed.
Measurement limit: 10 Hz – 10 KHz
Input voltage range: ≥ 0.6 V AC RMS (higher input voltage is required
at a higher frequency)
Input impedance: 10 MΩ
Maximum input voltage: 600 V AC RMS
Duty factor measurement (%)
Measuring range Resolution
Accuracy
0 – 100%
0.1%
±3%
In the duty factor measurement function:
Connect the red probe cable to the “INPUT” socket, and the black one
to the “COM” socket. Set the dial to the duty factor measurement func
-
tion (HZ%). Use the “Hz%” button to set the duty factor function (%).
Connect the measurement probes to the circuit to be measured. Note:
Do not exceed the input voltage, as this will damage the instrument.
If the duty factor is less than 10%, the “UL” symbol will appear on the
LCD display. If it is greater than 99.9%, the “UL” symbol will appear
on the LCD display.
Measurement limit: 1 Hz-10 MHz
Input voltage: ≥ 500 mV RMS