6.3 Manual Blood Pressure Measurement
Please note that the accuracy of the blood pressure depends from a valid calibration.
Furthermore it is in the responsibility of the physician to assess the absolute blood pressure values.
The prior attention of this method is the determination of the relative blood pressure changes caused
by other sleep related events.
It is necessary to perform a single Manual Blood Pressure Measurement.
The SpO2 Sensor and the ECG Electrodes must be applied to the Patient.
Make sure the recording has been started.
Apply the Blood Pressure Cuff from an external BP Monitor. Make a measurement and record the
results. At the time that the Blood Pressure Readings are obtained on the Blood Pressure Monitor,
mark this time point by pressing the Marker Button on the SOMNOwatch
™ plus.