9 Maintenance of
SOMNOwatch™ plus
9.1 Maintenance rate
Return the SOMNOwatch™ plus to SOMNOmedics after 3 years of usage for inspection. The
inspection includes Calibration of all the Recording Channels and visual inspection for any external
damage.
9.2 Cleaning and disinfection
Regularly clean the device to ensure trouble free operation.
Cleaning the device:
Wipe the
case
using only mild detergents. DO NOT use a wet or fluffy cloth.
Other disinfectants, approved by DGHM, may be used after testing that they do not damage the
surface of the device.
Do not use an autoclave for cleaning the
SOMNOwatch™ plus or any of its accessories.
Disinfection:
Article
Disinfectant
(trade mark)
Concentration Duration Intervals
Other
BP Adaptor Unit
Wipe disinfection:
For example:
Terralin Liquid
Mikrozid AF Cloth
Ready to use
Ready to use
5min
5 min
After use
Remove adhesives if
necessary. Follow
manufacturer’s instructions!
SpO2 Sensor
Wipe disinfection:
For example:
Terralin Liquid
Mikrozid AF Cloth
Ready to use
Ready to use
5min
5 min
After use
Remove adhesives if
necessary. Follow
manufacturer's instructions!
Elastic Belt
Quick disinfection
on alcohol basis:
For example:
Mikrozid AF
Liquid
Mikrozid AF Cloth
Isopropanol-
Spray
Ready to use
Ready to use
70%
5 min
5 min
5
–10 min
After use
Remove adhesives if
necessary. Follow
man
ufacturer’s instructions!
Use protective gloves when using aggressive disinfectants!
Follow the manufactures instructions when using disinfectants. Keep to the
prescribed dose and contact time!
Use cold water to prepare disinfection solvents!