showing that the transducer is set to 32 bits register size. Thus, by reading the third byte
of the response, you can deduce the value of Register 48.
c) Send the following message (Read Register 47):
Slave
Address
Command
Register
Address
Checksum
nn
03
00
47
10
00
02
checksum
where nn is the transducer's slave address.
The transducer will respond with either:
Slave
Address
Command
Data Bytes
Checksum
nn
03
04
4 data bytes
checksum
or:
Slave
Address
Command
Data Bytes
Checksum
nn
03
08
8 data bytes
checksum
Examine the first four bytes of the data. If they are all 00, then the transducer is in Big
Endian mode; if they are all FF, then the mode is Little Endian.
Summary of Contents for 7828
Page 1: ...Precision measurement solutions 7828 Density Transmitter TECHNICAL MANUAL 78285000 ...
Page 4: ...7828 Density Transmitter Technical Manual iv 78285000_AB ...
Page 10: ...7828 Density Transmitter Technical Manual x 78285000_AB ...
Page 60: ...Calibration check 7828 Density Transmitter Technical Manual 5 6 78285000_AA ...
Page 66: ...Maintenance 7828 Density Transmitter Technical Manual 6 6 78285000_AA ...
Page 72: ...Specification 7828 Density Transmitter Technical Manual A 6 78285000_AB ...
Page 80: ...Calculated parameters 7828 Density Transmitter Technical Manual B 8 78285000_AA ...
Page 82: ...7828 Density Transmitter Technical Manual Safety Certification C 2 78285000_AC ...
Page 106: ...7828 Density Transmitter Technical Manual COSHH Document F 2 78285000_AA ...
Page 108: ...7828 Density Transmitter Technical Manual COSHH Document F 4 78285000_AA ...
Page 109: ......