![Solartron Mobrey 7828 Technical Manual Download Page 78](http://html1.mh-extra.com/html/solartron-mobrey/7828/7828_technical-manual_1314071078.webp)
Calculated parameters
7828 Density Transmitter Technical Manual
B-6
78285000_AA
B.3 CALCULATED
PARAMETERS
These are also known as Special Functions
B.3.1 SPECIFIC
GRAVITY
Specific Gravity (SG)
Base density (@ Tref)
Density of water (@ Tref)
=
B.3.2 DEGREES
BAUMÉ
(Only available when Matrix referral is selected.)
Degrees Baumé = 145 -
145
Base density
where Base density is measured in g/cc.
B.3.3 DEGREES
BRIX
(Only available when Matrix referral is selected.)
Degrees Brix = 318.906 -
384.341
SG
SG
2
+
66 1086
.
where SG is Specific gravity.
B.3.4 QUADRATIC
EQUATION
(Only available when Matrix referral is selected.)
The following quadratic equation is implemented:-
y
A
B
d
C
d
B
B
= +
+
ρ
ρ
2
where:-
A, B ,C are User programmable constants.
d
= density of water (also a programmable constant).
ρ
B
= base density.
Summary of Contents for 7828
Page 1: ...Precision measurement solutions 7828 Density Transmitter TECHNICAL MANUAL 78285000 ...
Page 4: ...7828 Density Transmitter Technical Manual iv 78285000_AB ...
Page 10: ...7828 Density Transmitter Technical Manual x 78285000_AB ...
Page 60: ...Calibration check 7828 Density Transmitter Technical Manual 5 6 78285000_AA ...
Page 66: ...Maintenance 7828 Density Transmitter Technical Manual 6 6 78285000_AA ...
Page 72: ...Specification 7828 Density Transmitter Technical Manual A 6 78285000_AB ...
Page 80: ...Calculated parameters 7828 Density Transmitter Technical Manual B 8 78285000_AA ...
Page 82: ...7828 Density Transmitter Technical Manual Safety Certification C 2 78285000_AC ...
Page 106: ...7828 Density Transmitter Technical Manual COSHH Document F 2 78285000_AA ...
Page 108: ...7828 Density Transmitter Technical Manual COSHH Document F 4 78285000_AA ...
Page 109: ......