![Solartron Mobrey 7828 Technical Manual Download Page 27](http://html1.mh-extra.com/html/solartron-mobrey/7828/7828_technical-manual_1314071027.webp)
7828 Density Transmitter Technical Manual
Mechanical Installation
78285000_AB
2-13
2.3.6
FLOW-THROUGH CHAMBER INSTALLATION
This chamber is fabricated by Solartron, and is available with either weld prepared ends or
with flange or compression fittings for connection into the process pipe lines. It is available
with 1” or 2” inlet and outlet pipes.
Note:
The length of the inlet and outlet pipes must not be altered, otherwise the
temperature response and stability of the fitting may be adversely affected.
Conditions:
Flow:
constant, between 10 and 30 l/min
Viscosity limit:
Up to 20 000cP
Temperature:
-50 to +200
°
C
Pressure:
70 bar at 204°C, subject to process connections.
The PT100 is a direct insertion type, without a thermowell, and uses a ¾" Swagelok
connection.
The diagram below shows the dimensions of this type of standard installation.
A
B
Drain: ½”
1½”
¾”
Swagelok fittings
Flow in
Fl
ow o
u
t
Note: Datum lines must be preserved when
fitting flanges to weld-prepared tubes.
238mm (1” tubes) or
246mm (2” tubes)
Dims in mm.
A
B
1” Flange
480±3
335±3
1” Compr.
477±2
332±2
1” Weld
416±3
271±2
2” Flange
483±2
345±2
2” Compr.
522±2
384±2
2” Weld
409±2
271±2
1" Flow-through chamber
For optimum results the fluid flowing through the chamber should be homogeneous (plug
flow). This can be achieved by installing a static mixer immediately before the chamber.
Contact Solartron for advice about using flow-through chambers in hygienic applications.
Summary of Contents for 7828
Page 1: ...Precision measurement solutions 7828 Density Transmitter TECHNICAL MANUAL 78285000 ...
Page 4: ...7828 Density Transmitter Technical Manual iv 78285000_AB ...
Page 10: ...7828 Density Transmitter Technical Manual x 78285000_AB ...
Page 60: ...Calibration check 7828 Density Transmitter Technical Manual 5 6 78285000_AA ...
Page 66: ...Maintenance 7828 Density Transmitter Technical Manual 6 6 78285000_AA ...
Page 72: ...Specification 7828 Density Transmitter Technical Manual A 6 78285000_AB ...
Page 80: ...Calculated parameters 7828 Density Transmitter Technical Manual B 8 78285000_AA ...
Page 82: ...7828 Density Transmitter Technical Manual Safety Certification C 2 78285000_AC ...
Page 106: ...7828 Density Transmitter Technical Manual COSHH Document F 2 78285000_AA ...
Page 108: ...7828 Density Transmitter Technical Manual COSHH Document F 4 78285000_AA ...
Page 109: ......