78285000_AB
2-1
2
Mechanical Installation
2.1 INTRODUCTION
There are a variety of external factors which affect the ability of the 7828 to operate
successfully. In order to ensure that your system works correctly, the effects of these
factors must be taken into consideration when designing your installation.
There are two main aspects to consider:
1. The
accuracy and repeatability
of the measurements;
2. The
relevance
of the measurements to the
overall purpose
of the system.
Factors which may adversely affect
accuracy and repeatability
include:
•
=
The presence of gas or bubbles within the fluid being measured
•
=
Non-uniformity of the fluid
•
=
The presence of solids as contaminants
•
=
Fouling of the transducer
•
=
Temperature gradients
•
=
Cavitation and swirls caused by valves or discontinuities in the pipework
•
=
Operating at temperatures below the wax point of crude oils
•
=
The measured fluid being unrepresentative of the main flow.
In some applications, absolute accuracy is less important than repeatability. For example,
in a system where the control parameters are initially adjusted for optimum performance,
and thereafter only checked periodically.
The term
achievable accuracy
can be used to describe a measure of the product quality
that can be realistically obtained from a process system. It is a function of measurement
accuracy, stability and system response. High accuracy alone is no guarantee of good
product quality if the response time of the system is measured in tens of minutes, or if the
measurement bears little relevance to the operation of the system. Similarly, systems
which require constant calibration and maintenance cannot achieve good achievable
accuracy.
Factors which may adversely affect the
relevance
of the measurements could include:
•
=
Measurement used for control purposes being made too far away from the point of
control, so that the system cannot respond properly to changes;
•
=
Measurements made on fluid which is unrepresentative of the main flow.
Summary of Contents for 7828
Page 1: ...Precision measurement solutions 7828 Density Transmitter TECHNICAL MANUAL 78285000 ...
Page 4: ...7828 Density Transmitter Technical Manual iv 78285000_AB ...
Page 10: ...7828 Density Transmitter Technical Manual x 78285000_AB ...
Page 60: ...Calibration check 7828 Density Transmitter Technical Manual 5 6 78285000_AA ...
Page 66: ...Maintenance 7828 Density Transmitter Technical Manual 6 6 78285000_AA ...
Page 72: ...Specification 7828 Density Transmitter Technical Manual A 6 78285000_AB ...
Page 80: ...Calculated parameters 7828 Density Transmitter Technical Manual B 8 78285000_AA ...
Page 82: ...7828 Density Transmitter Technical Manual Safety Certification C 2 78285000_AC ...
Page 106: ...7828 Density Transmitter Technical Manual COSHH Document F 2 78285000_AA ...
Page 108: ...7828 Density Transmitter Technical Manual COSHH Document F 4 78285000_AA ...
Page 109: ......