![Solartron Mobrey 7828 Technical Manual Download Page 48](http://html1.mh-extra.com/html/solartron-mobrey/7828/7828_technical-manual_1314071048.webp)
Introduction
7828 Density Transmitter Technical Manual
4-6
78285000_AB
4.4.2 MENU
BAR
File
Exit
Another way of quitting ADView
Tools
Health Check
Determines whether the system is functioning
correctly
Register Read/Write
A facility for reading or writing to any of the
Modbus registers (see section 4.4.7)
Direct Comms
Enables you to specify exactly what will be
transmitted on the Serial link (see Appendix D).
Engineer Status
Only used by Solartron service engineers
Options
Simulate board
response/
Actual Board
Allows you to select between these two options
Enable / disable
screensaver
Allows you to select between these two options.
When enabled, the screensaver operates as
configured by the Windows system settings.
Window
Provides a means of opening or selecting
ADView’s facilities.
Help
Help on ADView
Limited help text, best read when first using
ADView.
About ADView
Displays software version number.
4.4.3
CONFIGURING A SLAVE ADDRESS
The 7828 factory configuration sets the slave address to 1. However, in many
applications it will be necessary to allocate another address. In a multi-drop application,
where several Modbus devices are connected on the same network, it is essential to
configure unique slave addresses for each device.
To do this, you will need to run ADView and use the Register Read/Write facility, detailed
in section 4.4.7. Check the value in Register 30 (Modbus Slave Address). If it is not the
required value, enter the desired value and click on the write button. The 7828 will now be
configured with the new slave address.
Summary of Contents for 7828
Page 1: ...Precision measurement solutions 7828 Density Transmitter TECHNICAL MANUAL 78285000 ...
Page 4: ...7828 Density Transmitter Technical Manual iv 78285000_AB ...
Page 10: ...7828 Density Transmitter Technical Manual x 78285000_AB ...
Page 60: ...Calibration check 7828 Density Transmitter Technical Manual 5 6 78285000_AA ...
Page 66: ...Maintenance 7828 Density Transmitter Technical Manual 6 6 78285000_AA ...
Page 72: ...Specification 7828 Density Transmitter Technical Manual A 6 78285000_AB ...
Page 80: ...Calculated parameters 7828 Density Transmitter Technical Manual B 8 78285000_AA ...
Page 82: ...7828 Density Transmitter Technical Manual Safety Certification C 2 78285000_AC ...
Page 106: ...7828 Density Transmitter Technical Manual COSHH Document F 2 78285000_AA ...
Page 108: ...7828 Density Transmitter Technical Manual COSHH Document F 4 78285000_AA ...
Page 109: ......