18
Setting Target Tolerances
This function will allow user to set plus and minus tolerances for torque and angle targets.
Note: These tolerances are used for manual modes only. Preset tolerances are defined by Minimum
and Maximum values.
1.
From Mode Setup menu, use
UP/DOWN
buttons to highlight tolerance selection to setup (TQ-%,
TQ+%, ANG-% ANG+%) then press
ENTER
button.
2.
Tolerance screen is displayed.
3.
Use
UP/DOWN
buttons to change tolerance value. Range is 0 to 10% (factory default for minus
tolerance is 0% and 4% for plus tolerance).
4.
Press
ENTER
button to accept selection and exit to Mode Setup menu.
Note: Green progress lights turn on at target minus -% TOL.
Note: Red progress lights turn on above target plus +% TOL.
Note: Plus tolerance is added to minimum Preset value to define initial maximum value when a Preset
is first added.
Enable/Disable Torque THEN Angle Mode
This function will allow user to enable or disable Torque THEN Mode.
1.
From Mode Setup menu, use
UP/DOWN
buttons to highlight
THEN DISABLED
(factory default)
selection then press
ENTER
button.
2.
TQ THEN ANGLE enable/disable screen is displayed.
3.
Use
UP/DOWN
buttons to select ENABLE or DISABLE selection.
4.
Press
ENTER
button to accept selection and exit to Mode Setup menu.
Note: Menu selection indicates current configuration (ENABLED or DISABLED).
Torque THEN Angle Mode
Torque THEN Angle mode is setup by first setting a target torque and units then a target angle before
selecting Torque THEN Angle mode. In Torque THEN Angle mode, when applied torque reaches target
torque, wrench automatically switches to angle mode for angle measurement. Progress lights indicate
applied torque progress while torque is measured and angle when angle is measured. If torque is
below target torque when angle reaches target angle, green progress lights will not turn on and if angle
exceeds maximum angle, red progress lights turn on indicating a potential problem with fastener.
ENTER
button
TORQUE -TOL%
0
ENTER
button
UP/DOWN
buttons
EXIT
TQ-% SETUP
TQ+% SETUP
TQ+% SETUP
ANG-% SETUP
ANG+% SETUP
ANGLE +TOL%
4
ENTER
button
UP/DOWN
buttons
ENTER
button
ENTER
button
UP/DOWN
buttons
EXIT
TQ-% SETUP
TQ+% SETUP
TQ-% SETUP
TQ+% SETUP
THEN DISABLED
TQ THEN ANGLE
ENABLE
DISABLE
TQ-% SETUP
TQ+% SETUP
THEN DISABLED
TQ THEN ANGLE
ENABLE
DISABLE
TQ-% SETUP
TQ+% SETUP
THEN ENABLED
Summary of Contents for ControlTech Micro
Page 35: ...34 ...