HOT LINE
®
Blood and Fluid Warmer
Technical Service Pack
49
S E C T I O N 5
E v e r y 1 2 M o n t h s
Every 12 Months
Alarm Testing
Place the HOTLINE
®
on a suitable firm surface. Visually check
that the recirculating solution level in the reservoir is above the MIN
mark. Attach a Disposable Set, ensuring that it is free from kinks and
twists.
Plug the unit into the electricity supply and switch on. Verify
that the unit is working normally (recirculating solution circulating,
temperature display gradually rising towards 41°C).
General Alarm Test
Press the General Alarm Test button. Observe:
■
the GREEN LED extinguishes.
■
the THREE RED LEDs light up.
■
the AUDIBLE ALARM commences beeping.
■
the RECIRCULATING SOLUTION ceases circulating.
If any of the above responses are missing, the unit must be
removed from service and repaired.