HOT LINE
®
Blood and Fluid Warmer
Technical Service Pack
57
S E C T I O N 5
M a i n t e n a n c e S o l u t i o n s
■
Switch off the HOTLINE® and dispose of the Disposable Set
in accordance with the usual procedures for your
establishment. Drain the Fluid from the HOTLINE® and
discard.
■
Refill the HOTLINE® with the second of the Isopropyl Alcohol
solution batches you made earlier.
NOTE:
If the HOTLINE
®
requires topping up at any time before the next
routine recirculating solution change, make sure you always use the
correct mix of Isopropyl Alcohol and Distilled water.
Changing Recirculating Solution (Hydrogen Peroxide)
At the annual routine service interval the Hydrogen Peroxide
solution must be replaced as follows:
■
Prepare two batches of 1.4 litres of recirculating solution
according to formula 1 above for 0.3% Hydrogen Peroxide
maintenance solution.
■
Drain the HOTLINE®, and refill with one of the prepared
batches.
■
Install an L-70 or L-70 NI Disposable Administration Set in
the HOTLINE®’s Disposable Set connector.
■
Turn on the HOTLINE® and allow the fresh fluid to circulate
for 30 minutes.
■
Switch off the HOTLINE® and dispose of the Disposable Set
in accordance with the usual procedures for your