HOT LINE
®
Blood and Fluid Warmer
Technical Service Pack
64
S E C T I O N 5
F u l l S e r v i c e T e s t P r o c e d u r e
Full Service Test Procedure
This sequence must be used:
■
either after any repair or service work where the case was
opened,
■
or when the unit has failed temperature calibration
verification.
Things you will need
■
Room Thermometer
■
TEMP CHECK HLTA-40
■
HOTLINE® Disposable Set L-70
■
Small and Medium Straight Screwdrivers
■
Non-magnetic Probe
CAUTION:
Take care while operating the HOTLINE
®
with the case front open.
Mains voltages are present inside. Avoid contact with any internal
components not specified in this procedure.
CAUTION:
The Auxiliary Outlet is for use only with approved accessories supplied
by your Smiths Medical distributor for that purpose.
CAUTION:
Do not fill the HOTLINE
®
reservoir with a Disposable Set in place.
Failure to remove the Disposable Set before the fill procedure may
result in an airlock in the HOTLINE
®
warmer.
NOTE:
When filling the HOTLINE
®
, use approved recirculating solutions
only. Failure to do so may cause damage to the equipment.
NOTE:
Check the calibration due date on the back of the TEMP CHECK before
commencing this procedure.
NOTE:
TEMP CHECK thermometers are designed for measuring circulating
recirculating solution flow only and are not intended for measuring
ambient (air) temperatures.
TEMPCHECK
GM-HL90-8502